IMT Medical Skip navigation links

 
aerisTM   Standard

 

The aeris  medical air compressor provides high-quality compressed air for a wide variety of medical devices, including adult and infant ventilators. 

 

The combination of high flow performance with exceptional low-cost and reliability allows the aeris to stand apart from other medical air compressors on the market. User selectable output pressure, a central air pipeline backup system and RS-232 interface make the aeris a compressor you can depend on.

 

Approvals: CE, CSA, FDA

Buy Online 

 Technical Specifications 

Pneumatic Data Basic Standard High Flow
Output flow (continous) 45 L/min at 3 bar (43 psi) 45 L/min at 3 bar (43 psi) 55 L/min at 3 bar (43 psi)
Output flow (peak) 250 L/min 300 L/min 300 L/min
Internal tank size 0.2 L 2 L 2 L
Air quality - Medical grade according
to ISO 8573
Medical grade according
to ISO 8573
Oil and particulate filtration 5 µm (micron) 0.01 µm (micron) 0.01 µm (micron)
Active carbone filtration no yes yes
Dew point depression > 3°C below room temperature
at peak output pressure
> 10°C below room temperature
at peak output pressure
with optional membrane dryer
> 3°C below room temperature
at peak output pressure
> 10°C below room temperature
at peak output pressure
with optional membrane dryer
> 3°C below room temperature
at peak output pressure
> 10°C below room temperature
at peak output pressure
with optional membrane dryer
User Configurable Functions
Output pressure adj. 1 to 3 bar (14 to 43 psi)
adj. 1 to 4 bar (14 to 57 psi)
with optional membrane dryer
adj. 1 to 3 bar (14 to 43 psi)
adj. 1 to 4 bar (14 to 57 psi)
with optional membrane dryer
adj. 1 to 3 bar (14 to 43 psi)
adj. 1 to 4 bar (14 to 57 psi)
with optional membrane dryer
Adjustable pressure
threshold for automatic backup
not available 1 to 8 bar (14 to 114 psi) 1 to 8 bar (14 to 114 psi)
Physical Data
Width, depth, height of device 35 x 35 x 55 cm 35 x 35 x 55 cm 35 x 35 x 55 cm
Weight of device 31 kg (67 lbs) 34 kg (74 lbs) 34 kg (74 lbs)
AC power 100 VAC (50/60Hz)
115 VAC (50/60Hz)
230 VAC (50/60Hz)
100 VAC (50/60Hz)
115 VAC (50/60Hz)
230 VAC (50/60Hz)
100 VAC (50/60Hz)
115 VAC (50/60Hz)
230 VAC (50/60Hz)
Power consumption 520 VA 520 VA 520 VA
Noise below 50 dBa at 1m below 48 dBa at 1m below 48 dBa at 1m
Compliance & Approvals

CE marking, EN60601-1 (safety, EN60601-1-2 (EMC), EN737, EN739,

FDA (only aeris Standard with optional membrane dryer)

Environmental Data
Operating temperature 10 to 40°C (50 to 104°F) 10 to 40°C (50 to 104°F) 10 to 40°C (50 to 104°F)
Humidity below 85% relative humadity
below 95% relative humidity
with optional membrane dryer
below 85% relative humadity
below 95% relative humidity
with optional membrane dryer
below 85% relative humadity
below 95% relative humidity
with optional membrane dryer