IMT Medical Skip navigation links Skip navigation links

MultiGasAnalyserTM   OR-703

 

imtmedical is expanding the capabilities of the FlowAnalyser with the MultiGasAnalyser OR-703 by offering instant gas concentration measurements of CO2, N2O, Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane.


The imtmedical MultiGasAnalyser OR-703 probe is designed using the latest advances in micro system technology and provides a complete in line real time monitoring system with unique versatility and design. The sensor proves his reliability when testing the performance of any anesthesia delivery and monitoring systems or the accuracy of CO2 monitoring devices. The MultiGasAnalyser OR-703 sensor head measures infrared light absorption at several different wavelengths and exactly determines the gas concentrations of the mixtures.

 Technical Specifications 

Measurements
CO2 0 - 10%
N2O 0 - 100%
HAL, ISO, ENF 0 - 5%
SEV 0 - 8%
DES 0 - 18%
Response Time
CO2 < 60ms
N2O and the
5 aneasthetic agents
< 150ms
Monitoring
Interface RS-232 port
Numerical data available through the FlowAnalyser.
Numerical data and real-time curves available through FlowLab software.
Physical Data
Weight < 30 g (exl. cable)
Size 3.7 x 2.7 x 3.5 cm (1.45 x 1.1 x 0.9 inches)
Internal battery yes
Dimensions (w x d x h) 22 x 25 x 12 cm
Weight 3.7 kg
Environmental
Operating 10 - 40 °C
Storage -20 - 50 °C
Humidity 10 - 95% r.H., non-condensing
Atm. pressure 700 - 1200 mbar (3048 m)
Compliance and Approvals
CE marked according to the 93/42/EEC MDD
ISO 11196:1997, EN 864:1996, EN 12598:1000, ISO/DIN 21647:2003,
ISO 7767, ASTM-F 1462-92, ASTM-F 1456-92 and ASTM-F 1462-93

 

Measurement Tolerance

This table does not contain any measurement tolerance information. More details in these brochures.