FlowAnalyserTM PF-300
|
|
|
|
|
Measuring your pneumatic equipment for reliability and precision with a dependable calibration tool is critical in avoiding fatal errors. Precision and reliability are exactly what the FlowAnalyser Product Line provides.
The FlowAnalyser measures flow, pressure, temperature, humidity and O2 concentrations bidirectionally. The one-of-a-kind Adult, Pediatric and High Frequency ventilation measuring modes make the FlowAnalyser the ideal calibration tool for all ventilators, anaesthesia machines and spirometers.
The FlowAnalyser distinguishes itself from other calibration tools by combining a simple, intuitive multilingual user interface with the highest precision. With the push of a button, all measured values can be stored directly on the FlowAnalyser and later retrieved for documentation purposes. FlowLab software complements the FlowAnalyser by offering a wide range of graphical analysis capabilities. Swiss devotion to detail at its best.
|
|
|
|
|
|
|
| Flow & Pressure Measurements |
|
PF-300 |
PF-301 |
PF-302 |
| Flow |
Measuring direction |
bidirectional |
. |
. |
. |
|
Temperature compensated |
yes |
. |
. |
. |
| Pressure compensated |
yes |
. |
. |
. |
| Humidity compensated |
yes |
. |
. |
. |
| O2 compensated |
yes |
. |
. |
. |
| High |
± 300 L/min * |
. |
. |
. |
| Low |
± 20 L/min * |
. |
. |
. |
| Pressure |
High |
0-10 bar * |
. |
. |
. |
|
Average |
± 150 mbar * |
Diff |
Rel |
Rel |
| Low |
0-5 mbar * |
|
|
. |
| High in Flow Canal |
0-150 mbar * |
. |
. |
. |
| Barometer |
0-1150 mbar (abs) * |
. |
. |
. |
| Vacuum pressure |
± 1000 mbar * |
|
. |
|
| Measuring Units |
Flow |
L/min, L/s, cfm, mL/min, mL/s |
. |
. |
. |
|
Pressure |
bar, mbar, cmH2O, inH2O, Torr,
inHg, hPa, kPa, mmHg, PSI |
. |
. |
. |
|
Additional Measuring Values |
| Oxygen |
Concentration |
0-100% |
. |
. |
. |
|
Pressure compensated |
yes |
. |
. |
. |
| Temperature |
High in Flow Canal |
0-50°C |
. |
. |
. |
| Dew Point |
High in Flow Canal |
-10-50°C |
. |
. |
. |
| Humidity |
High in Flow Canal |
0-100% |
. |
. |
. |
| CO2 |
Concentration |
0-10% ** |
with OR-703 |
| N2O |
Concentration |
0-100% ** |
with OR-703 |
| HAL, ISO, ENF |
Concentration |
0-12% ** |
with OR-703 |
| SEV |
Concentration |
0-15% ** |
with OR-703 |
| DES |
Concentration |
0-22% ** |
with OR-703 |
| Gas Types |
|
Air, Air/O2, N2O/O2, Heliox (21% O2), He/O2, N2, CO2, customized gas types |
. |
. |
. |
| Gas Standardisation |
|
ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPD,
0/1013, 20/981, 15/1013, 25/991, 20/1013 |
. |
. |
. |
|
Respiratory Parameters |
| Rate |
|
1-1000 bpm |
. |
. |
. |
| Time |
TI, TE |
0.05-60 s |
. |
. |
. |
| I:E ratio |
|
1:300 - 300:1 |
. |
. |
. |
| Ti/Ttotal |
|
0-100% |
. |
. |
. |
| Breath Volumes |
Vti, Vte |
±10L |
. |
. |
. |
| Minute Volumes |
Vi, Ve |
0-300 L/min |
. |
. |
. |
| Pressure |
Ppeak, Pmean, PEEP, Pplateau |
0-150 mbar |
. |
. |
. |
| Peakflow |
Peakflow Insp./Exp. |
±300 L/min |
. |
. |
. |
| Compliance |
Cstat |
0-1000 mL/mbar |
. |
. |
. |
| Trigger |
Adult, Pediatric, HFO |
Adjustable on flow or pressure
curves with user-definded limits. |
. |
. |
. |
|
General Information |
| Electrical & Physical Data |
AC input |
90-260 VAC, 50/60 Hz |
. |
. |
. |
|
Battery (lead recharcheable battery) |
3 hrs (with OR-703 2 hrs) |
. |
. |
. |
| Power consumption |
23 VA |
. |
. |
. |
| Weight |
3.7 kg |
. |
. |
. |
| Dimensions (w x d x h) |
22 x 25 x 12 cm |
. |
. |
. |
| Data Storage |
|
all parameters (measured as well as respiratory values) |
. |
. |
. |
| Display |
Graphic display |
Intuitive user interface with numerical
measuring values, statistics, volume trigger
configuration, gas type selection
and calibration menus. |
. |
. |
. |
| RT-200 Emulation Mode |
|
Simulates RT-200 style commands over the RS-232 interface. |
. |
. |
. |
| Communication Interfaces |
|
USB for Windows Software FlowLab, RS-232 for individual communication, TTL for external trigger. |
. |
. |
. |
| Calibration |
|
annually |
. |
. |
. |
| Conditions |
Ambient temperature |
10-40°C (50-104°F) |
. |
. |
. |
|
Humidity |
10-95% R.H. |
. |
. |
. |
| Approvals |
|
CE, CSA |
. |
. |
. |
|
|
Measurement Tolerance |
|
* These specifications do not contain any measurement tolerance information. More details in these brochures.
** These specifications do not contain any measurement tolerance information. More details in these brochures. |
|
|