For our most sensitive patients


Highest precision in volume and pressure – and a unique user experience.

Reliability and safety are key in ventilation. One of the most demanding challenges is to ventilate neonates, where you need highest precision in volume and pressure. bellavista combines latest technology with precise delivery of pressure or volume for our smallest patients. bellavistas software design supports adding new features, simply by updates. So all in all, this leads to a safe, complete and future proof ventilator with a unique user experience.

Mehr Informationen

Invasive Ventilation


TargetVent

Newborn patients ventilated with volume targeted modes are more likely to survive without lung damage. Ventilator assistance is needed for a shorter duration and target controlled ventilation helps to prevent pneumothorax and leads also to more stable carbon dioxide levels. bellavistas neonatal Target Vent provides a full ventilation platform with target volumes for extremely low birthweight patients, up from 2 mL of tidal volume. Our TargetVent Modes can also be used for Non-Invasive Ventilation.

neonatal target vent

Non-Invasive ventilation


Target Vent

Target controlled ventilation for pressure supported, spontaneous and non-invasive modes are possible, because of our precise and fast auto.leak leakage compensation and our proximal iFlow 40 flow sensor, which scans with a very high sampling rate for changes in ventilation flow patterns. So you have more ventilation options at hand with bellavista at any time.


nCPAP and nIPPV

The bellavista nCPAP can be applied as flow or pressure based mode. The nCPAP Mode features a breath rate detection and gives you additional safety through apnoea alarming.

With our biphasic non-invasive mode nIPPV, you can operate with high frequencies up to 200 bpm.

bellavista screen neonatal suite
bellavista neonatal suite

Burst Backup

Our Burst Backup feature gives you maximum safety in nCPAP ventilation. With an adjustable number of breaths you can prevent apnoea without changing the mode to apnoea ventilation. So the neonate will keep its own spontaneous breathing rhythm.

Sensitive Triggering


To detect spontaneous triggering signals from neonates you need absolute precise sensors. Our advanced proximal iFlow 40 S Sensor is strongly build and detects tiniest and finest signals with a high sampling rate.

Oxygen


Neonatal O2 Flush Customisable O2 flush capability allows the user to set the maximum of FiO2 required and the duration of delivery.

Weaning PSVtarget


Our mode that can be used for convenient weaning with a pressure regulated volume guaranteed PSVtarget and adaptive pressure support.

Technical specification Neonatal Suite


Parameter Specification
bellavista 1000 Neonatal Suite
Patient Types Neonatal
Technology High performance turbine
Ventilation Modes
  • Pressure controlled
CPAP, PSV, P-A/C, PC-SIMV
  • Volume controlled
P-ACtarget, PC-SIMVtarget, PSVtarget
  • Non Invasive Modes
CPAP, PSV, P-AC, PC-SIMV, P-ACtarget, PC-SIMVtarget, PSVtarget, nCPAP, nIPPV, HFOT
  • Adaptive Mode
-
  • Apnea Ventilation
P-AC, PC-SIMV, nIPPV
  • Backup Modes
PSV, Burst backup
Inspiratory pressure 2..45 mbar
Psupport 2..45 mbar
PEEP 0..20 mbar
Tidal volume 2 ml..250 ml (with TargetVent option)
Inspiratory time 0.1..2 sec
Respiratory rate 0..150 bpm
I:E ratio 1:59; 5:1
Inspiratory Trigger Flow 0.1..20 l/min, Pressure 0.1..15 mbar, Trigger off
Expiratory Trigger 5..90%
Rise time auto.rise, 0..400 ms manual
Leakage Compensation auto.leak, full In- and Expiratory compensation
Curves Pressure, Flow, Volume, SpO2, etcCO2
Loops Pressure Volume, Pressure Flow, Flow Volume
Monitoring 45 online parameters
Maneuvers Inspiratory Hold, Expiratory Hold, NIF(Negative Inspiratory Force), Vtrapped, P0.1 (Occlusion Pressure), AutoPEEP, Manual Breath
O2 Flush FiO2 and duration configurable
Graphics TargetVent view
Trending Monitoring Parameters and Curves
Oxygen 21% to 100%, high and low pressure oxygen inlet
Options Sidestream Capnography
SpO2 Plethysmography
Patient circuit types Single Limb and Dual Limb Circuits
Dimensions (w × h × d) 350 × 220 × 330 mm
Screen 13.3" Colour, full Touchscreen operation
Battery time max. 6h (internal)
Interfaces 2 × RS232, 2 × USB, Can Bus, Ethernet, CO2, SpO2, Nurse Call, VGA
Weight 12.8 kg
Power input 100-240 VAC/50-60 Hz, low voltage input 24 VDC/3.5 A

Interested in bellavista?